P-E-02—Procedure for generating and issuing calibration console certificates for consoles compliant with the requirements of EL-ENG-12-01 and S-E-01

Category: Electricity
Procedure: P-E-02 (rev. 2)
Issue Date: 2013-07-01
Effective Date: 2013-07-01
Supersedes: P-E-02 (rev.1)
Documents: EL-ENG-12-01, S-E-01


Table of content


1.0 Purpose

This document describes the procedures for generating and issuing Calibration Console Certificates for consoles that have been tested and found compliant to the requirements of EL-ENG-12-01—Requirements for the Certification of Measuring Apparatus—Electricity Meter Calibration Consoles and S-E-01—Specifications for the Calibration, Certification and Use of Electricity Calibration Consoles.

2.0 Scope

This document is applicable to Calibration Console Certificates issued pursuant to the requirements of EL-ENG-12-01 and S-E-01.

3.0 References

3.1 EL-ENG-12-01– Requirements for the Certification and Use of Measuring Apparatus—Electricity Meter Calibration Consoles

3.2 P-E-01—Procedures for Calibrating and Certifying Calibration Consoles Pursuant to the Requirements of EL-ENG-12-01

3.3 E-17—Automatic Renewal of Certificates for Electricity Calibration Consoles

3.4 S-E-01—Specifications for the Calibration, Certification and Use of Electricity Calibration Consoles

4.0 Background

The roles and responsibilities for issuing Calibration Console Certificates ensuing from the calibration of consoles are set forth in section 5.2. of EL-ENG-12-01 and S-E-01. This document provides procedures, notes and templates that will aid in the generation of certificates. It also identifies requirements for maintaining and updating certificates.

5.0 Policy

5.1 Procedures for generating calibration console certificates

5.1.1 New certificates

  1. Ensure that all applicable testing has been completed and documented in the P-E-01 worksheets.
  2. Ensure that the worksheets indicate who completed the testing and the date the tests were completed.
  3. Enter all data as required on the first page of the certificate using the sample certificate found in Annex A as a template.
  4. Assign a unique certificate number in accordance with the following numbering system:

    E-AAA-BB-CCC

    Where:

    • E = Electricity
    • AAA = EST, ONT and WES for MC Eastern, Ontario and Western region, respectively, or a 3 letter designation of a C-D-01-delegated organization
    • BB = last two digits of the issue year of certificate
    • CCC = sequential number assignment (001-999)
  5. Assign an expiry date that is one year from the date the tests were completed.
  6. Modify the last paragraph of the sample certificate as required so that it reflects only those quantities certified for use with the console.
  7. Enter any bias transformer settings that may be applicable to the console in the "Additional Notes, Comments and Restrictions" section.
  8. Enter the appropriate revision number(s) if the console uses any software or firmware.
  9. Enter all reference meter information as applicable on the certificate.
  10. Enter any restrictions or other notes concerning the certification or use of the console.
  11. Complete the tables of errors for each quantity certified for use with the console.
  12. Ensure that table header information is included on each new page that contains table data.
  13. Delete, add, or change the headings of columns as required to accommodate design, operation and use specific to the console (see examples in Annex A).
  14. Enter the minus (-) sign for negative errors.
  15. Ensure that each page is numbered indicating the total number of pages of the certificate. The certificate number should be placed on each page of the certificate.
  16. Include a column for uncertainty error values, or provide a separate table, depending on what is applicable (see examples).
  17. Generate a cover letter using the sample provided in Annex A as a template.
  18. Ensure that the certificate and cover letter are signed by the responsible authority delegated to certify calibration consoles and sent to the owner of the console together with a copy of the completed worksheets.
  19. Send a copy of the certificate, worksheets and covering letter to the local Regional Electricity Specialist (the issuing authority must retain a copy).

5.1.2 Extension certificates

  1. Ensure that the console meets the criteria for annual renewal of certificates as outlined in bulletin E-17.
  2. Issue an extension certificate similar to the original certificate; however, revise the certificate number to include the suffix identifying the renewal, as per section 6 of E-17.
  3. Ensure that the new expiry date reflects the one-year extension.
  4. Send the extension certificate to the owner of the console with an attached cover letter (see sample cover letter in Annex A). One copy is to be sent to the local Regional Electricity Specialist, and one copy retained by the issuing authority.

5.1.3 Amendments

  1. Calibration Console Certificates that require corrections, additions, or other amendments may be reissued under the original certificate number; however, a revision number shall be included with the certificate number:

    E-AAA-BB-CCC rev.xx

    Where:

    • E-AAA-BB-CCC = Original certificate number
    • rev.xx = Revision number xx, where "xx" is sequentially ordered starting with 01.

6.0 Revisions

6.1 The purpose of this revision is to update the procedure in accordance with the revision made to the electricity meter calibration console requirements in EL-ENG-12-01.

6.2 The purpose of revision 1 was to update the requirements in accordance with the changes made in S-E-01.

Annex A—Example documents

A.1 Sample of a completed certificate—without official error tables (issued following certification test)

Sample 1 page 1

The Measurement Canada federal identity program bilingual corporate signature is placed in the upper-left corner. The English-first version is used in English documents and the French-first version is used in French documents.

Certificate No.: E-WES-12-001
Date of test:
Expiry Date:

Calibration Console Certificate

Owner: A.B.C. Public Utility
Manufacturer: Console Makers Ltd.
Serial No.: 123456
Location: Anywhere, Alberta

Model: 0001
No. of Positions: 10 Energy, 10 Demand
Owner's No.: 890

This is to certify that the above-stated calibration console was examined, duly calibrated with respect to reference standards, and found to comply with all applicable requirements outlined in sections 7 and 8 of the Electricity and Gas Inspection Regulations and EL-ENG-12-01. A statement of the calibration console errors is attached.

The calibration console is certified to be used as a measuring apparatus for the purpose of testing electricity meters as specified hereunder.

This calibration console is hereby certified to be used for the verification, reverification and sample testing of single-phase and poly-phase watt hour, volt ampere hour, and var hour energy meters and watt, volt ampere, and var demand meters.

Signature appears here

Delegated Authority

Date appears here

Date

Page 1 of 3

The "Canada" wordmark is placed in the bottom-right corner.

Sample 1 page 2

Console Owner: A.B.C. Public Utility
Console No.: E-WES-12-001
Date:

Reference Meter Data

Energy Reference Meter

  • Type (Wh, VAh, Varh): Wh, Varh
  • Manufacturer: XYZ
  • Model: 4000
  • Serial No.: 123456
  • Pulses rate: 100,000 pulses/unit
  • Type (Wh, VAh, Varh): VAh (AVG & RMS)
  • Manufacturer: XYZ
  • Model: 4000
  • Serial No.: 123457
  • Pulses rate: 100,000 pulses/unit
  • Type: Volts, Amps
  • Manufacturer: XYZ Standards
  • Model: ABC
  • Serial No.: 123458
  • Pulses rate: N/A

Demand Reference Meter

  • Type (W, va, var): Watt, Var
  • Manufacturer: ABC Consoles
  • Model: Internal Standard 3
  • Serial No.: 123459
  • Response type: RMS
  • Type (W, va, var): Watt, VA
  • Manufacturer: ABC Consoles
  • Model: Internal Standard 4
  • Serial No.: 123460
  • Response type: AVG, RMS
  • Type (W, va, var): Var
  • Manufacturer: ABC Standards
  • Model: XYZ
  • Serial No.: 123461
  • Response type: RMS

Additional Notes, Comments or Restrictions

 

Page 2 of 3

Sample 1 page 3

Console Owner: A.B.C. Public Utility
Console No.: E-WES-12-001
Date:

Reference Meter Information

Serial No.: 123456
Model: 4000
Owner's No.: 890

Quantity being measured (W, Wh, VA, Vah, etc.): W and Wh

Are connections parallel or multiple? multiple

Are there 1:1 transformers in the circuit? ­­­­Yes

Are there intervening transformers in the test circuit? No

What is the highest Uc(cons) for energy? 0.04% for k (coverage) factor of 2

What is the highest Uc(cons) for demand? 0.04% for k (coverage) factor of 2

Official Console Errors
Voltage Current Phase
angle
Voltage
tap
Current
tap
Selector switch setting MUT position Official console error
120 2.5 120 5 2.5 1 0.03
120 2.5 60° 120 5 2.5 1 -0.04
120 50 120 100 50 1 0.00
120 50 60° 120 100 50 1 -0.01
240 2.5 240 5 2.5 1 -0.11
240 2.5 60° 240 5 2.5 1 0.09
240 50 240 100 50 1 0.08
240 50 60° 240 100 50 All 0.00

Page 3 of 3

A.2 Sample cover letter

Sample 2

The Measurement Canada federal identity program bilingual corporate signature is placed in the upper-left corner. The English-first version is used in English documents and the French-first version is used in French documents.

Standards Building
151 Tunney's Pasture Drive
Ottawa, Ontario
K1A 0C9

Édifice des normes
151, promenade Tunney's Pasture
Ottawa (Ontario)
K1A 0C9

A Console Owner
234 Main Street
Mississauga, ON
A1B 2C3

Attention: Mr. Kent

Dear Mr. Kent:

Re: Calibration Console Certificate

Please find enclosed Calibration Console Certificate E-ONT-12-013 for your calibration console # E1234.

Sincerely,

Paul Special
Electricity Specialist Ontario Region
paul.special@canada.ca

cc Regional Manager, Ontario Region

Enclosure

The "Canada" wordmark is placed in the bottom-right corner.

A.3 Sample extension cover letter

Sample 3

The Measurement Canada federal identity program bilingual corporate signature is placed in the upper-left corner. The English-first version is used in English documents and the French-first version is used in French documents.

Standards Building
151 Tunney's Pasture Drive
Ottawa, Ontario
K1A 0C9

Édifice des normes
151, promenade Tunney's Pasture
Ottawa (Ontario)
K1A 0C9

A Console Owner
234 Main Street
Mississauga, ON
A1B 2C3

Attention: Mr. Kent

Dear Mr. Kent:

Re: Calibration Console Certificate

Please find enclosed Calibration Console Certificate E-ONT-12-013-2 for your calibration console # E1234. The certificate has been extended to .

At the term of the second extension, the calibration console must be calibrated against the full range of tests currently specified in EL-ENG-12-01. If the calibration console meets all the applicable requirements, it will be certified by the Regional Electricity Specialist for a period of one year.

For more information, please contact the local Measurement Canada Electricity Specialist.

Sincerely,

Ray Special
Electricity Specialist Ontario Region
ray.special@canada.ca

cc Regional Manager, Ontario Region

Enclosure